ANALYSEUR DE BIOCHIMIE BS-200E
- Discrete, random access, fully automated, bench-top.
- Constant 200 tests per hour, up to 330 tests per hour with ISE.
- Auto washing system with pre-warmed detergent and water.
- Grating optical system.
- 150?l minimum reaction volume.
- Independent mixing bar.
- Reusable reaction cuvettes.
- Bi-directional connection to LIS host.
Référence : G127019
État : Neuf
System Functions:
- Automatic, Discrete, Random Access.
- STAT sample priority.
Throughput: Constant 200 tests/hour (without ISE), up to 330 tests/hour with ISE.
Principles: Absorbance photometry, Turbidimetry, Ion Selective Electrode technology.
Methodology: End-point, Fixed-time, Kinetic, optional ISE Single/Dual reagent chemistries, monochromatic/bichromatic Linear/non-linear multi-point calibration.
Programming: Open system with user defined profiles and chemistry calculation System pack reagents ready to use.
Reagent/Sample Handling:
Reagent/Sample Tray: 40 reagent positions, 40 sample positions in cooling compartment (2~12°C).
Reagent Volume:
- R1: 10~350ul, step by 1 ul.
- R2: 10~200ul, step by 1 ul.
- Sample volume: 2~45ul, step by 0.1 ul.
Reagent/Sample Probe:
- Liquid level detection and tracking, vertical & horizontal collision protection and inventory checking
- Probe cleaning: Automatic washing of interior and exterior
- Carry-over < 0.1%.
Automatic sample dilution:
- Pre-dilution and post-dilution.
- Dilution ratio up to 1: 200.
Internal Bar Code Reader (optional):
- Used for sample and reagent scan.
- Applicable to various bar code systems such as
- Codabar, ITF (Interleaved Two of Five), code128, code39, UPC/EAN, Code93.
- Bi-directional interface LIS transmission.
ISE Module (optional): Measure K+, Na+, Cl?.
Optical System:
- Light Source: Halogen-tungsten lamp.
- Photometer: Grating system, reversed optics.
- Wavelength: 12 wavelengths, 340nm, 380nm, 412nm, 450nm, 505nm, 546nm, 570nm, 605nm, 660nm, 700nm, 740nm and 800nm.
- Absorption range: 0~3.3Abs (10mm conversion).
- Resolution: 0.0001Abs.
Reaction System:
- Reaction rotor: Rotating tray, containing 80 cuvettes.
- Cuvette: Reusable, optical length 5mm.
- Reaction volume: 150~500ul.
- Reaction temperature: 37°C.
- Temperature fluctuation: ±0.1°C.
Mixing System: Standalone mixing bar.
Cuvette Washing: 8-step washing station with pre-heated detergent and water.
Control and Calibration:
- Calibration mode: Linear (one-point, two-point and multi-point), Logit-Log 4P, Logit-Log 5P, Spline,
- Exponential, Polynomial, Parabola Control software: Westgard multi-rule, Cumulative sum, check, Twin plot, L-J Chart.
Operation Unit:
- Operation system: Windows® XP Professional/Home SP2 or above Windows® 7.
- Interface: RS-232.
Working Conditions:
- Power Supply: AC 200~240V, 50/60Hz, <=1500VA or AC 100~130V, 50/60Hz, <=1500VA.
- Temperature: 15-30°C for operation.
- Humidity: 35-85% RH.
Dimension: 860mm (W) x700mm (D) x625mm (H).
Weight: 130 Kg.