Principles : Flow Cytometry (FCM), Semi-conductor Laser scatter, chemical dye method, independent Basophil channel Impedance method for WBC, RBC, PLT counting Cyanide-free reagent for Hemoglobin test.
Parameters
27: WBC, Lym%, Mon%, Neu%, Eos%, Bas%, Lym#, Mon#, Neu#, Eos#, Bas#, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, LIC%, LIC#, ALY%, ALY# 3 histograms and 1 scattergram.
Throughput : Up to 60 samples per hour.
Sample
- Volume Whole Blood: 20 ?L.
- Prediluted: 20 ?L.
Test Mode: CBC CBC+DIFF.
Performance
Carryover Precision Linearity
WBC? 0.5% ? 2.0% (4-15 x 10?/L) 0.00-99.99 x10?/L
RBC? 0.5% ? 1.5% (3.5-6.0 x 10¹²/L) 0.00-8.00 x10¹²/L
HGB? 0.5% ? 1.5% (110-180g/L) 0-250g/L
PLT? 1.0% ? 4.0% (150-500 x 10?/L) 0-1000 x10?/L
Data Storage Capacity : Up to 40,000 results including numeric and graphical information.
Communication : LAN Port supports HL7 protocol.
Operating Environment
- Temperature: 15°C~30°C.
- Humidity: 30~85%.
- Air Pressure: 70~106 kPa.
Power Requirement : A.C 100-240V ? 300VA 50/60Hz.
Dimension and Weight
- 410mm(L) x 470mm(W) x 530mm(H).
- Weight: ?45 Kg.