Principles
Impedance method for WBC, RBC and PLT counting Cyanide free reagent for hemoglobin test.
Performance
Parameter Linearity Range Precision (CV %) Carryover WBC
(109 /L) 0-100 ?3.5% (4.0-6.9 ?0.5?
?2.0% (7.0-15.0)
RBC(1012/L) 0-8.00 ?1.5% (3.5-6.5) ?0.5?
HGB(g/L) 0-280 ?1.5% (100-180) ?0.5?
MCV(fL) ?1.0% (70-110)
PLT(109 /L) 0-1000 ?5.0% (100-149) ?1.0?
?4.0% (150-500)
Parameters
19 parameters: WBC, Lymph#, Mid#, Gran#, Lymph%, Mid%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT 3 histograms for WBC, RBC and PLT.
Reagent
- M-30D DILUENT.
- M-30CFL LYSE.
- PROBE CLEANSER.
Sample
- Volume Prediluted mode 20?L.
- Whole blood mode 9 ?L.
Throughput: 40 samples per hour.
Display : 8.4 inch TFT Touch Screen.
Data Storage Capacity : Up to 200,000 results including numeric and graphical information.
Communication
- LAN Port supports HL7 protocol.
- Support bi-directional LIS.
Interface: 4 USB port (for external printer, software upgrade, barcode reader, WIFI adapter,keyboard and mouse), LAN port (1).
Printout :Thermal recorder, 50 mm wide paper, various printouts formats External printer optional.
Operating
- Environment Temperature: 10°C~40°C.
- Humidity: 10%~90%.
- Air pressure: 70kPa?106kPa.
Power Requirement
- 100V-240V.
- ?300VA.
- 50Hz/60Hz.
Dimension and Weight Dimension: Depth(410 mm) x width(300 mm) x height(400 mm) Weight: ?20Kg.